摘要
By removing the disturbance of determined arsenic with the atomic fluorescent spectrometry,which by copper ion and other metal ion in copper concentrate and copper ore,to build a suited dissolution method,and to choose the optimization condition,to overcome the shortage of general atomic fluorescent spectrometry dissolved the sample.The detection limit of the method is 1μg/g.This method is applied to three national standard samples(level one) to detect the content of arsenic.The detected values and recommended ones are consistent.Therefore,the method is feasible to test the content of arsenic in high copper concentrate samples.